Single Beam Visible Spectrophotometer LVS-101

Single Beam Visible Spectrophotometer 100 Series from Labocon features auto calibrating spectrophotometer with visible spectrum measuring range having diffraction grating optical system for optimum photometric performance.
Specifications:
Main Specification:
Wavelength Repeatability:
0.5 nm
Wavelength Accuracy:
±1 nm
Wavelength Range:
340-1000 nm
Spectral Bandwidth:
4 nm
Photometric Range:
0-100%T
0-3.000A
Photometric Accuracy:
±0.5%T
Photometric Repeatability:
0.2%T
Stray Light:
≤0.2%T at 360 nm
Light Source:
Tungsten Halogen Lamp
Detector:
Silicon Photodiode
Dimension:
510×490×280 mm
Weight:
6.5 kg