Home > Products > Visible Spectrophotometer > Vis Spectrophotometer LVS-A10

Vis Spectrophotometer LVS-A10

Labtron manufactures LVS-A10 having single beam optical system. It uses silicon photo diode detector and tungsten lamp as light source which provides the best wavelength coverage across the working range of the spectrometer.

Wavelength Range

335~1000nm

Setting Wavelength

332~1003nm

Spectral Bandwidth

5 nm

Wavelength Accuracy

±2 nm

Wavelength Repeatability

1nm

Photometric Range

T:0~199.9% , A:-0.3~2.5Abs, F:0~9999, C:0~9999

Photometric Accuracy

±1.0%T

Photometric Repeatability

0.5%T

Detector

Silicon photo diode

Light Source

Tungsten lamp (Socket Type, 20W / 12V, 2000h)

Stray Light

≤0.5%T(360nm)

Stability

≤0.5%T/5min

Power Requirement

220V/50Hz

Dimensions

385×330×185mm

Weight

10kg

Calibration: Automatic 0%(T) and 100%(T)
Display of concentration data function
1200 lines/mm grating
High accuracy with digital wavelength display

  1. Quantitative and qualitative analysis, detection of Impurities and functional groups determination.